Detail Memmert Incubator Laboratorium I IF 75
The incubator I is perfect for the world of research, medicine, pharmaceutics and food analytics, as well as food chemistry.
Setting temperature range : +20 to +80 °C
Working temperature range : min. 10°C above ambient up to +80°C
Setting accuracy temperature : 0.1 °C
Temperature sensor : 1 Pt100 sensor DIN class A in 4-wire-circuit
Language setting : German, English, Spanish, French, Polish, Czech, Hungarian
ControlCOCKPIT SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display
Timer : Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT : the process time does not start until the set temperature is reached
Calibration : three freely selectable temperature values
adjustable parameter : temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime
Fan forced air circulation by quiet air turbine, adjustable in 10 % steps
Fresh air : Admixture of pre-heated fresh air by electronically adjustable air flap
Vent : vent connection with restrictor flap
Documentation ; programme stored in case of power failure
Programming : AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand).
Temperature control adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Autodiagnostic system : for fault analysis.